electric iv syringe pump

Item No.: YSZS-1800C
Features:
1.Stackable to be the maximum 8 channels.
2.Compatible with disposable syringes
  from 5ml,10ml,20ml,30ml,50ml,60ml,universal type.
3.High accuracy +/-2% after correct calibrated.
4.Automatically detect the remained solution.
5.Three work
  • BrandYsenmedical
  • Model numberYSZS-1800C
  • Warranty1 year
  • Customizedyes
  • Lead time7 days
  • Packagecarton
  • Description
    YSZS-1800C electric iv syringe pump
    Features:
    1. Stackable to be the maximum 8 channels.
    2. Compatible with disposable syringes
       from 5ml,10ml,20ml,30ml,50ml,60ml,universal type.
    1. High accuracy +/-2% after correct calibrated.
    2. Automatically detect the remained solution.
    3. Three work modes: Rate mode, Time mode and
    Weight mode
    1. Unique LCD displayer.
    2. Bolus / Purge, KVO function.
    8.  12 Hours continuous worked.
     
     
    Technical Specifications
    PARAMETER SPECIFICATION
    Applicable syringes 5ml,10ml,20ml,30ml,50ml,60ml of any standard
    Accuracy +/- 2% with correct calibration
    Rate mode 50ml(60ml)0.1ml/h-1800.0ml/h (0.1ml/h step)
    20ml(30ml)0.1ml/h-800.0ml/h (0.1ml/h step)
    10ml0.1ml/h-400.0ml/h (0.1ml/h step)
    5ml: 200ml/h
    Time Mode SOL. VOL:0.1-999.9ml (0.1ml step)
    TIME1-1999min (1min step)
    Body Weight Mode: Volume0.1-999.9ml(0.1ml step)
    DRUG MASS0.1-999.9mg 0.1mg step)
    DOSE0.01-99.99ug/kg/min(0.01ug/kg/min step)
    WEIGHT0.1-300kg(0.1kg the least step)
    Total Volume Range 0.1-1999.9ml (0.1 ml step)
    Purge Rate 50ml(60ml)1800.0ml/h,
    20ml(30ml)800.0ml/h,
    10ml400.0ml/h, 5ml: 200ml/h
    KVO Rate 1ml/h
    Occlusion pressure Three level setting LOW, MEDIUM or HIGH
    Alarms Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures.
    Three levels of alarm volume LOW, MEDIUM or HIGH
    Bolus Rate 5ml:10~200ml/h,10ml:10-300ml/h,20ml(30ml):10-600ml/h,
    50ml(60ml):10-1200ml/h
    Bolus Volume Maximum of 25ml
    Interface RS-232 Reserved port
    Power source AC power supply: AC 100 to 240V( +/-10%)50/60Hz;
    Built-in Lithium Battery:  Rechargeable, 11.1V, 2000mAh.
    12 hours of continuous operation.
    Accessories: AC power line, Operational manual
    Protection Class Class I, Type CF,IPX4
    Dimension / Net weight 230mm*150mm*125mm , 2.0kg
     
     


    Technical Specification:

    Infusion rate range
    Adjustable drop rate range: 1 /min~400 drops/min 
                              (step: 1 drop /min)
    Adjustable flow rate range: 1ml/h~1200ml/h
                             (step 0.1ml/h)
    Purge rate: 800ml/h
    KVO rate: 1ml/h (less than 100ml/h)
              2ml/h (100~300ml/h)
              3ml/h (greater than 300ml/h)
    Infusion volume range: 1ml~9999.9ml (step: 0.1 ml)
    Maximum accumulated volume: 9999.9ml (step: 0.1 ml)
    Infusion Time range: 1min~9999min (step: 1 min)

    Infusion accuracy: Within ±3% (using calibrated high quality IV set)
                       Mechanical precision: within ±2%

    Occlusion alarm threshold:
             High: 800 mmHg ± 200 mmHg (106.7kPa±26.7kPa)
    Medium: 500 mmHg ± 100 mmHg (66.7kPa±13.3kPa)
    Low: 300 mmHg ± 100 mmHg (40.7kPa±13.3kPa)
    Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle.

    Size: 515mm(L)×145mm(W), allowed deviation: ±5%

    SpO2 sensor
    Type: Finger clip type
    Wire length: 3000mm, allowed deviation: ±5%
    Human body contact materials: Silicone

    Size: 188 mm(L)×198 mm(W)×228 mm(H)
    Net weight: 2.5kg

    Technical Specification:

    Infusion rate range
    Adjustable drop rate range: 1 /min~400 drops/min 
                              (step: 1 drop /min)
    Adjustable flow rate range: 1ml/h~1200ml/h
                             (step 0.1ml/h)
    Purge rate: 800ml/h
    KVO rate: 1ml/h (less than 100ml/h)
              2ml/h (100~300ml/h)
              3ml/h (greater than 300ml/h)
    Infusion volume range: 1ml~9999.9ml (step: 0.1 ml)
    Maximum accumulated volume: 9999.9ml (step: 0.1 ml)
    Infusion Time range: 1min~9999min (step: 1 min)

    Infusion accuracy: Within ±3% (using calibrated high quality IV set)
                       Mechanical precision: within ±2%

    Occlusion alarm threshold:
             High: 800 mmHg ± 200 mmHg (106.7kPa±26.7kPa)
    Medium: 500 mmHg ± 100 mmHg (66.7kPa±13.3kPa)
    Low: 300 mmHg ± 100 mmHg (40.7kPa±13.3kPa)
    Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle.

    Size: 515mm(L)×145mm(W), allowed deviation: ±5%

    SpO2 sensor
    Type: Finger clip type
    Wire length: 3000mm, allowed deviation: ±5%
    Human body contact materials: Silicone

    Size: 188 mm(L)×198 mm(W)×228 mm(H)
    Net weight: 2.5kg